Measurement Functions | AC/DC Withstanding Voltage & IR |
Interface | RS232/Handler/USB |
Arc Detection | 9 Levels (AC/DC) |
Other Functions | LIST Edit Discharge Function 5“ LCD + Touch panel |
AC Withstand Voltage | ● |
Rate Ouput | AC 10.000kV/20mA |
Output Voltage | 0.05kV-10.000KV AC Resolution: 10V |
Voltage Accuracy | 2% of reading + 5 digits |
Output Regulation | 2% of reading +5V |
Output Frequency | 50Hz/60Hz |
Measurement Current | 0.001mA-20.00mA Resolution: 0.001 / 0.01 |
Current Accuracy | 2% of reading + 5 digits |
Rising Time | 0.4s-999.9s |
Testing Time | 0.5s-999.9s |
Falling Time | 0.1s-999.9s |
DC Withstand Voltage | ● |
Rate Output | DC 10kV/10mA |
Output Voltage | 0.05kV-10.000KV DC Resolution: 10V |
Voltage Accuracy | 2% of reading + 5 digits |
Output Regulation | 2% of reading +5V |
Current Measurement | 0.1μA-10.000mA |
Current Accuracy | 2% of reading +5 digits |
Rising Time | 0.4s-999.9s |
Testing Time | 0.5s-999.9s |
Falling Time | 0.1s-999.9s |
Insulation Resistance | ● |
Rate Output | 2500Vdc/50GΩ |
Output Voltage | 0.05kV-2.500kV DC Resolution: 1V |
Voltage Accuracy | 2% of the set value + 5 V |
Power Supply | 110VAC~220VAC 47Hz~63Hz 300VA |
Dimension & Weight | 479mm( W )x133mm( H )x391mm ( D ) 15kg |
Accessories | ATL900E Withstand voltage test line ATL108 RS232 Communication DB9-handler |